Electric Conductivity Cell for X series
Product Name | Immersion type conductivity cell "CALMEMO" ![]() |
Immersion type conductivity cell | |||
Model | For general use CT-58101B |
For low electric conductivity CT-58101C |
For high electric conductivity CT-58101A |
For general use C-50101B |
|
dimensional drawing | ![]() |
![]() |
![]() |
![]() |
|
measurement range | electric conductivity | 100μS/m〜10S/m {1μS/cm〜100mS/cm} |
5μS/m〜1S/m {0.05μS/cm〜10mS/cm} |
1mS/m〜100S/m {10μS/cm〜1S/cm} |
100μS/m〜10S/m {1μS/cm〜100mS/cm} |
temperature | 0〜100℃ | 0〜100℃ | 0〜100℃ | ─ | |
cell constant | 100m-1 | 10m-1 | 1000m-1 | 100m-1 | |
required amount of specimen * | 4mL (Φ14mm×26mm or more) |
6mL (Φ14mm×42mm or more) |
39mL (Φ36mm×38mm or more) |
4mL (Φ14mm×26mm or more) |
|
temperature sensor | ○ | ○ | ○ | ─ | |
remarks | JP17 designated cell The cell constants were determined using standard solution of potassium chloride (1408μS/cm at 25℃) according to JIS K 0130 . |
JP17 designated cell The cell constants were determined using standard solution of potassium chloride (146.5μS/cm at 25℃) according to JIS K 0130 . |
─ | ─ |
Product Name | flow-type electric conductivity cell "CALMEMO" ![]() |
||||
Model | for general use CT-88101B |
for low electric conductivity CT-88101C |
for high electric conductivity CT-88102A |
for purified water CT-27111D (flow cell is not included) |
|
dimensional drawing | ![]() |
![]() |
![]() |
![]() |
|
measurement range | electric conductivity | 100μS/m〜10S/m {1μS/cm〜100mS/cm} |
5μS/m〜1S/m {0.05μS/cm〜10mS/cm} |
10mS/m〜100S/m {100μS/cm〜1S/cm} |
5μS/m〜20mS/m {0.05μS/cm〜200μS/cm} |
temperature | 0〜100℃ | 0〜100℃ | 0〜100℃ | 0〜80℃ | |
cell constant | 100m-1 | 10m-1 | 2000m-1 | 1m-1 | |
temperature sensor | ○ | ○ | ○ | ○ | |
Remarks | 4mL required in cell | 15mL required in cell. It is possible to measure low electric conductivity as there is no effect from carbon dioxide. |
1mL required in cell. Flow cell for a small amount. Dropper rubber and pump tube are required separately. |
8mL required in flow cell. Flow cell is required separately. PP made (CEF-22A) (connecting port diameter: outside diameter 8mm × internal diameter 4mm) SUS made (CEF-23A) (connecting port diameter: outside diameter 8mm × internal diameter 6mm) |
![]() |
"CALMEMO": The sensor itself stores the model and serial number. As cell constant is also stored in memory, it reads the cell constant automatically when you connect the cell to the main unit. |
- * The capacity when using the smallest container that can be inserted into a cell. Do not make a suddenly change of temperature of 50℃ or more.